Development of a national protocol for a high activity (192)Ir brachytherapy source calibration.

Development of a national protocol for a high activity (192)Ir brachytherapy source calibration. Research Abstract Details 

Research Abstract Table of Contents

Jump to the:

  • Abstract Text of This Paper
  • Journal Published
  • MeSH Keywords of This Abstract
  • Chemicals and Substances Used in this Paper
  • Grants and Granting Agency of this Research
  • Database Accession Numbers Used in this Paper
  • Related Papers
  • Related Research Tags
  • Rate this Research Paper
  • Development of a national protocol for a high activity (192)Ir brachytherapy source calibration. Abstract Text:

    n t milosevi?N T Milosevi?,j stankovi?J Stankovi?,n t milosevi?N T Milosevi?,j stankovi?J Stankovi?,

    Purpose: Determination of the air kinetic energy released in material (KERMA) rate (commonly named source strength) or a brachytherapy source activity represents one of the first steps in the acceptance, as well as in the periodic Quality Assurance (QA) testing procedures. In this paper, calibration of an (192)Ir high-dose rate (HDR) brachytherapy source is presented. Materials and methods: An (192)Ir HDR brachytherapy source calibration was carried out, with two independent procedures - calibration in water and in free air. The dosimeter used consisted of a thimble ionization chamber (NE 2571) and an electrometer (NE 2670). The reference air KERMA rate was calculated according to the International Atomic Energy Agency (IAEA) and the International Commission for Radiation Units and Measurements (ICRU) protocols. All physical quantities and correction factors used were determined, calculated, and re-checked for particular chamber and source type. Results: In water, the reference air KERMA rate, determined by the IAEA protocol differed by +1.0% from the manufacturer's declared value and was measured with a relative error of 1%. The reference air KERMA rate, calculated by the ICRU protocol differed by -0.4% from the manufacturer's declared value and was measured with a relative error of 1.1%. In air, the value of the reference air KERMA rate, calculated by the IAEA protocol, differed by +1.6% from the manufacturer's declared value and was measured with a relative error of 1%. The value of the reference air KERMA rate for the same source, calculated by the ICRU protocol, differed by +1.1% from the manufacturer's declared value and was measured with a relative error of 0.9%. The results showed that application of the IAEA protocol gave higher values than the application of the ICRU protocol. It stands for both calibration methods. The difference between calibration protocols was not higher than 2% (in the water 1.4% and in the air 0.5%). Finally, calibration results showed that the difference between the calibration methods (in water and in air), for both calibration protocols, was not higher than 2%. For the IAEA protocol the difference was -0.6% and for the ICRU protocol it was -1.5%. We believe that benefits can be derived from the simultaneous application of both methods. Conclusion: A proposal of a national protocol represents an attempt to formulate a unique protocol for brachytherapy sources calibration in Serbia. The comparison of "in water" and "in free air" calibration has shown systematic differences of in-source strength of about 2%, respectively. It would be worth suggesting to radiotherapy centres to adopt the most optimal calibration method, based on well-known international protocols.

    Development of a national protocol for a high activity (192)Ir brachytherapy source calibration. Publishing Authors By Initials

    nt milosevi?NT Milosevi?,j stankovi?J Stankovi?,nt milosevi?NT Milosevi?,j stankovi?J Stankovi?,

    For similar abstracts research abstracts see: abstracts research

    PUBMED ID PMID:

    MEDLINE DATE: 2002 Jul-Sep

    Development of a national protocol for a high activity (192)Ir brachytherapy source calibration. Journal Published:

    PUBLICATION TYPE: Journal Article

    Journal: Journal of B.U.ON. : official journal of the Balka

    VOLUME: 7

    Page Numbers: 261-71

    Journal Abbreviation:

    ISSN: 1107-0625

    DAY: 8

    MONTH: 10

    YEAR: 2007

    Development of a national protocol for a high activity (192)Ir brachytherapy source calibration. Information

    Number of References:

    LANGUAGE: eng

    NlmUniqueID: 100883428

    Development of a national protocol for a high activity (192)Ir brachytherapy source calibration. Keywords Mesh Terms:

    KEYWORDS:

    MESH TERMS:

    Chemical & Substance for Abstract: Development of a national protocol for a high activity (192)Ir brachytherapy source calibration. Information

    Substance Name:

    Registry Number:

    Grant and Affiliation Information for Development of a national protocol for a high activity (192)Ir brachytherapy source calibration.

    AFFILIATION: Institute of Biophysics, Medical School, University of Beograd, Beograd, Yugoslavia.

    Country: Greece

    Greece Research PublicationGreece Research Publication

    AGENCY:

    GRANT:

    ACRONYM:

    MEDLINETA: J BUON

    REFSOURCE:

    DATABASENAME:

    ACCESSION NUMBER:

    Number Hits: 0

    Development of a national protocol for a high activity 192Ir brachytherapy source calibration Related Publications

     

    Molecular Station USER Menu

    Welcome to Molecular Station!

    You have to register before you can post on our forums or use our advanced features. Register Now! Its Free and Fast!

    Already registered? Login now below.

    User Name:

    Password:

    Already registered and Forgot your password? Click below to recover it.

    Recover Lost Password

    Join now - it's fast and free!

    Molecular Station is THE largest network of researchers, scientists and science lovers anywhere!

    Research Terms of Usage and Disclaimer
    Home
    Features

    Protocols

    DNA Forum

    Science Forum

    DNA Forum
    Biology Forum

    Science News


    [CaRP] XML error: Invalid document end at line 2

    For more click here:Science News