BACKGROUND: Procalcitonin (PCT) is a biomarker for the diagnosis of sepsis and bacterial infection diseases. METHODS: A new fully automated SphereLight PCT (SL-PCT) assay system for PCT concentration in human serum or plasma by using SphereLight 180 (SL180, Olympus Corp.) analyzer was developed. The SL-PCT assay is based on chemiluminescent enzyme immunoassay. RESULTS: A linear dose response relationship was observed up to 200 ng/ml PCT concentration. The detection limit of PCT concentration was 0.06 ng/ml. Endogenous substances, anticoagulants, sodium fluoride and drugs did not interfere with assay results. There was a good correlation between the present method and the manual method in serum and plasma samples. CONCLUSIONS: These results indicate that the SL-PCT assay showed good performance in terms of the linearity, detection limit and precision. Use of this PCT measurement may improve the detection of sepsis and infectious disease.