Calibration of Becton Dickinson Flow Cytometers for Relative Fluorescence Intensity Measurements - Details

Link Details 

Protocols » Listing Details
ID:1918
Title: Calibration of Becton Dickinson Flow Cytometers for Relative Fluorescence Intensity Measurements - http://cyto.mednet.ucla.edu/Protocols/relative.htm
Description:Flow cytometers must be calibrated prior to fluorescence intensity measurements because of inherent instrument variability. To correct for this variability, a standard particle (fixed chicken red blood cells, or CRBCs) must be analyzed on the instrument prior to each experiment and photomultiplier tube (PMT) voltages adjusted accordingly to place the CRBC fluorescence emission peaks into predetermined target channels.
Category:Cell Biology and Cell Culture: Flow Cytometry Protocols: Flow Cytometry Calibration Protocols