Warning: ob_start() [ref.outcontrol]: output handler 'ob_gzhandler' conflicts with 'zlib output compression' in /home/molec2/public_html/protein/protein-concentration-protocol/index.php on line 2
Protein Concentration Protocols

Bioinformatics, Protocols, DNA RNA Protein Proteomics

Sponsor / Advertise | Link to us | Contact us | About us | Help us

home > protein > protein-concentration-protocol > index.php

tlw tlw2

Welcome to Molecular Station!

You have to register before you can post on our forums or use our advanced features. Register Now! Its Free and Fast!

Already registered? Login now below.

User Name:

Password:

Already registered and Forgot your password? Click below to recover it.

Recover Lost Password

Join now - it's fast and free!

Molecular Station is THE largest network of researchers, scientists and science lovers anywhere!

Molecular Biology - Science Quotes

The great men of science are supreme artists. ~Martin H. Fischer

Molecular Biology Newsletter!

Yes! I Want to Learn the Latest in Molecular Biology and Research! Please Make Me an Expert in My Lab Work!
Also I Want to Tell My Friends to Get My Free PCR Chapter Please! 
Don't Worry Your Email is Safe with Us. We hate Spam as Much as You Do. 
First Name:
Email:

Recent Forum Posts

 

Protein Determination by UV Absorption Protocol

Protein Determination by UV Absorption Protocol, ( Modified from the protocol by Alastair Aitken and Michèle P. Learmonth)

Materials for UV Protein Concentration

1. 0.1 M K2SO4 (pH 7.0).
2. 5 mM potassium phosphate buffer, pH 7.0.
3. Nonionic detergent (0.01% Brij 35)
4. Guanidinium-HCl.
5. 0.2-µm Millipore (Watford, UK) filter.
6. UV-visible spectrometer: The hydrogen lamp should be selected for maximum     intensity
at the particular wavelength.
7. Cuvets, quartz, for <215 nm.

Protein Concentration UV Method

 Estimation of Protein by Near UV Absorbance (280 nm):

1. A reliable spectrophotometer is necessary. The protein solution must be diluted in the
buffer to a concentration that is well within the accurate range of the instrument (see Notes 1 and 2).

2. The protein solution to be measured can be in a wide range of buffers, so it is usually no problem to find one that is appropriate for the protein which may already be in a particular buffer required for a purification step or assay for enzyme activity, for example (see Notes 3 and 4).

3. Measure the absorbance of the protein solution at 280 nm, using quartz cuvets or cuvets that are known to be transparent to this wavelength, filled with a volume of solution sufficient to cover the aperture through which the light beam passes.

4. The value obtained will depend on the path length of the cuvet. If not 1 cm, it must be
adjusted by the appropriate factor. The Beer-Lambert law states that: A (absorbance) = ε c l where ε = extinction coefficient, c = concentration in mol/L and l = optical path length in cm. Therefore, if ε is known, measurement of A gives the concentration directly, ε is
normally quoted for a 1-cm path length.

5. The actual value of UV absorbance for a given protein must be determined by some absolute method, e.g., calculated from the amino acid composition, which can be determined by amino acid analysis. The UV absorbance for a protein is then calculated according to the following formula: A280 (1 mg/mL) = (5690nw + 1280ny + 120nc)/M
where nw, ny, and nc are the numbers of Trp, Tyr, and Cys residues in the polypeptide of
mass M and 5690, 1280 and 120 are the respective extinction coefficients for these residues (see Note 5).

See also Protein Concentration Protocols from other labs.

 

 

Other Protein Articles:

Western Blot Home

Immunoprecipitation Home

SDS-PAGE Gel Electrophoresis

Protein Separation

Protein Purification

Protein Identification

Recombinant Protein Expression Systems

Protein Expression Techniques

 

 

Bid, Buy and Sell on eBay Disclaimer / Terms of Service | Privacy Policy| ©2005-2007 Molecular Station.com, All rights reserved.

send to a friend Send this page to a friend